Języcki:
pl en cz
Katalog produktów

Cement Retained PreFormance® Post Single Or Multi-Unit

Cement Retained PreFormance® Post
Single Or Multi-Unit

 undefined

ABUTMENT SELECTION
5.
Measure the tissue depth in the interproximal areas at each location and select the proper abutment collar height that will allow the margin to be at tissue level or slightly below after preparation. Also, select the desired emergence profile and a straight or 15º angled post. Finally, match the color of the implant analog to determine the platform diameter.

undefined 

6. Place the selected PreFormance Posts into each analog. Line up the hexes and place the flat side of the post to the buccal. Press firmly until feeling and/or hearing the audible and tactile click. Secure the abutments into place using a Certain® Titanium Abutment Screw and the Large Hex Driver.

undefined 

7. Place the vacuum formed template on the master cast over the PreFormance Posts. Reduce the posts as necessary so these fit within the template. Block out the undercuts on the adjacent teeth. Place the cast back on the articulator. Apply a separator to the posts and the casts. Fill the tooth portion of the template with acrylic resin. Fully seat the template on the cast over the abutments using the articulation. Allow the acrylic to set per the manufacturer’s instructions.


undefined 

PROVISIONAL FABRICATION
8.
Place the vacuum formed template on the master cast over the PreFormance Posts. Reduce the posts as necessary so these fit within the template. Block out the undercuts on the adjacent teeth. Place the cast back on the articulator. Apply a separator to the posts and the casts. Fill the tooth portion of the template with acrylic resin. Fully seat the template on the cast over the abutments using the articulation. Allow the acrylic to set per the manufacturer’s instructions.

or

Duplicate the master cast with the abutments in place. Articulate, wax the provisional on the duplicate cast and process it in acrylic resin.

undefined 

9. Remove the template from the PreFormance Posts. Remove the provisional from the template. Remove all excess acrylic from around the margin areas of the provisional. Fill in any voids. Finish the crown or bridge to the desired contour and polish.

OPTIONAL: Relieve each abutment area for the intraoral reline of the provisional.

undefined 

CT GUIDED SURGICAL IMPLANT PLACEMENT
10.
CLINICIAN
Place the implants using the surgical guide and following the surgical plan provided by the guide manufacturer.

See page 8 for an example of a surgical plan.


undefined 

POST SURGICAL DELIVERY OF PROVISIONAL
11.
Place each PreFormance Post into the implants, one by one, following the tooth position numbers on the buccal. Press firmly until feeling and/or hearing the audible and tactile click. Thread the titanium abutment screw into each implant using the Large Hex Driver. Verify an accurate fit of each abutment by visualizing the interfaces or take a radiograph. Torque the abutment screws to 20Ncm using the Large Hex Driver Tip and a torque device. Seal each access hole with a temporary filling material. Try in the provisional over the PreFormance Posts and verify it fits to the margins properly. Adjust the occlusion as indicated and remove the provisional.

undefined 

12. If the provisional did not fit passively, place acrylic resin into each abutment area, seat the provisional onto the abutments and have the patient close into occlusion. Allow the acrylic to set per the manufacturer’s instructions. Remove the provisional. Fill in any voids. It may be necessary to remove the PreFormance Posts and place those into the provisional to fill marginal voids. Remove any excess acrylic and polish. Place temporary cement into the provisional, seat it on the abutments and have the patient close into occlusion. Remove any excess cement from around the margin areas. Allow the cement to set per the manufacturer’s instructions. If a flap procedure was used during surgery, suture the tissue around the PreFormance Posts and the provisional.